Programming
248083389-1 Rev
7-5
Setting the Real Time Clock
The Real Time Clock (RTC) can be read in registers 40027 and 40028:
Note:
The Real Time
Clock can also be set on
screen by accessing the
CONFIG screen and
selecting the Clock button.
In order to change the RTC to the current local date/time, enter the high
and low values as unsigned integers to registers 40035 and 40036
respectively, which are the Data Set registers.
Then write the command
13
to the command register 40002. This will
write the values in the Data Set registers (40035 and 40036) to the RTC
registers (40027 and 40028).
Changing the Default Instrument Parameters
The main instrument parameters involved with the operation of the
REMOTE counter are Location, Sample Time, Hold Time and Initial
Delay.
The Location is set by writing an unsigned integer to register 40026.
The range of values is from 0 to 999.
Sample Time, Hold Time and Initial Delay all use 2 registers, a high
word and a low word. If the desired value for any of these parameters
is less than 9 hours, then only the low word register needs to be written
(in seconds).
Table 7-2 Real Time Clock Registers
Register
Data Type
Description
40027
signed integer
Real Time Clock (RTC) [high]. Works in
conjunction with 40028. Displays date and
time, in number of seconds since
midnight, 1/1/1970.
40028
signed integer
Real Time Clock [low]
Table 7-3 Data Set Registers
Register
Data Type
Description
40035
unsigned integer
Data Set [high]. Works in conjunction
with 40036. Data entered here is applied
to the device through the command
register.
40036
unsigned integer
Data Set [low]
Содержание solair 3100
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Страница 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Страница 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Страница 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Страница 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Страница 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Страница 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Страница 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Страница 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Страница 187: ...00...
Страница 188: ......