Operate the ApexPortable
248083440-1 Rev 5
5-49
•
When "REPORT" is enabled, the "LOC" and "SAVE" buttons are
displayed only when there are at least the minimum number of
locations needed for the recipe area available.
•
If REPORT is not active or the number of locations do not meet the
minimum needed in setup, the LOC and SAVE buttons are not
displayed. See Figure 5-79.
Figure 5-79 Report Setup Screen - No Locations
The next section contains an example of setting up the instrument to
run a report.
Set Up Counter to Run Reports
For this example, the FED STD FT
3
report will be used. See Figure 5-
80.
Figure 5-80 FED STD FT3 Report Screen
•
Set the instrument for the Fed Std ft
3
report, 100 ft
2
, Class 100,
unidirectional air flow, In Operation room status.
•
This will generate the following minimum requirements:
•
Min Locations: 2
•
Min Samples: 5
Содержание ApexP3
Страница 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Страница 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Страница 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Страница 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Страница 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Страница 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Страница 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Страница 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Страница 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Страница 129: ...00...