ApexPortable MODBUS Register Map v1.50
248083440-1 Rev 5
A-5
Device Status
The Device Status registers (40003 and 40057) display the current
status of the device (Table A-4). Additional status bits are shown in
40056 (Table A-5).
.
40065
signed integer
Run-time particle channel alarm low flags (bit 0 = channel 1,
…).
40074
signed integer
ApexPortable
:
Last Calibration Date [high]. Indicates when instrument was
last calibrated. This number can be generated by the ANSI C/
C++ time() function.
40075
signed integer
ApexPortable
:
Last Calibration Date [low]
Table A-3 Sensor Settings Registers
Register Data
Type
Description
Table A-4 Current Device Status (40003/40057)
Bit
Description
0
RUNNING:
Set when a start command is executed via
Command 11 (instrument start) or through the user
interface. The flag will remained set until a stop
command is executed.
1
SAMPLING:
This is set only when the instrument is
actually sampling data that is to be recorded. Caution
must be used in sending a command during this time that
may invalidate current sample.
2
NEW DATA:
Set to 1 to indicate that a new data record
has been recorded and it hasn't been read via modbus yet.
When a data record has been read via modbus (registers
30001 to 30999), then this flag is reset to zero.
3
DEVICE ERROR:
in the event that there is a failure on
the device, this bit is set to indicate possible invalid data
has been collected.
11
ApexPortable:
LASER STATUS:
Set to 1 when unit’s LASER is out of
spec, else set to 0.
Содержание ApexP3
Страница 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Страница 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Страница 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Страница 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Страница 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Страница 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Страница 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Страница 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Страница 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Страница 129: ...00...