Operate the ApexPortable
248083440-1 Rev 5
5-45
•
Use the Report Setup screen as shown in Figure 5-69.
Figure 5-69 Report Setup Screen
•
Press the CLASS button and the Select Class screen displays as
shown in Figure 5-70. Touch the desired class so it becomes
highlighted. Press the BACK button to save and return to previous
screen.
Figure 5-70 Select Class Level
•
The selected level will be highlighted in blue.
Class Levels vary based on which standard has been selected in the
PRINT REPORT field. Below are the available Class Levels, per
standard:
FED_209E ft
3
- 1, 10, 100, 1000, 10000, 100000
ISO14644-1 - 1, 2, 3, 4, 5, 6, 7, 8, 9
EU GMP - A, B, C, D (Rev 2009)
Changing the Class Level will also change the minimum number of
locations, minimum number of samples and the minimum sample
volume per channel size required to satisfy the requirements of the
chosen report.
Содержание ApexP3
Страница 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Страница 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Страница 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Страница 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Страница 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Страница 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Страница 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Страница 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Страница 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Страница 129: ...00...