Operate the ApexPortable
248083440-1 Rev 5
5-11
DATA SETUP
Figure 5-14 shows the Data Setup option buttons. These options allow
the user to configure the parameters for data collection.
Figure 5-14 Data Setup Options
PARTICLE
The instrument’s particle channels can be enabled or disabled on the
Particle screen. A checkmark is displayed next to each enabled
channel. See Figure 5-15.
Figure 5-15 Particle Channel Configuration
•
The button on the right of each channel size is an ON/OFF toggle.
Press once to toggle a channel between a
, disabling the channel
and the default
, enabling the channel.
When channels are disabled, they are removed from the MAIN screen
display, from the reports and from the printouts.
Disabled
Enabled
Press button to
Change State
then
Enable the
particle
sizes to
be measured
Содержание ApexP3
Страница 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Страница 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Страница 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Страница 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Страница 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Страница 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Страница 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Страница 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Страница 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Страница 129: ...00...