Battery is very Low:
battery icon at the lower left
corner of the screen: battery charge insufficient for
testing.
Meter will not operate and battery icon disappears
when the test strip is removed. Replace the battery.
Battery Icon
Troubleshooting
CAUTION: Attempts to perform
glucose testing with a low battery
could result in the software rebooting.
Upon rebooting, the software version
is displayed for 3 seconds. Software
version is numeric (example 5.0);
therefore, please exercise caution
to ensure the software version is not
reported as a glucose result.