2. Disinfect the Meter.
• Using a new, fresh germicidal bleach
wipe, thoroughly wipe the surface of
the meter (top, bottom, left, and right
sides) a minimum of 3 times horizon-
tally followed by 3 times vertically
avoiding the electrical connector.
• Gently wipe the surface area of the test strip port
making sure that no fluid enters the port.
3. Observe surface contact time.
• Ensure the meter surface stays wet for
1 minute and is allowed to air dry for an
additional 1 minute.
Cleaning and Disinfecting the Meter