Interfering Substances
Glucose Interferences:
The OneTouch StatStrip Xpress Glucose Hospital Meter Sys-
tem exhibits
no
interference from the following list of drugs
or physiological conditions commonly found in patients in
intensive care settings with complex medical conditions and
drug regiments up to the following concentration levels:
Tested
Interfering Substances
Tested
Concentration Level
Acetaminophen
20.0 mg/dL
1.323 mmol/L
Acetoacetate
51.0 mg/dL
5 mmol/L
Acetone
69.7 mg/dL
12 mmol/L
Acyclovir
0.6 mg/dL
0.027 mmol/L
Albuterol
0.06 mg/dL
0.0025 mmol/L
Amitriptyline
0.06 mg/dL
0.0022 mmol/L
Amoxicillin
5 mg/dL
0.137 mmol/L
Ampicillin
0.8 mg/dL
0.0229 mmol/L