Clinical Study Performance
Table A-1 Arterial Specimens
Analytical and System Accuracy Comparison: Glucose con-
centrations < 75 mg/dL
Site
Within
± 5 mg/dL
Within
±10mg/dL
Within
±12mg/dL
Within
±15mg/dL
Exceeds
±15mg/dL
Site #1
64/82
75/82
78/82
79/82
3/82
(78.0%)
(91.5%)
(95.1%)
(96.3%)
(3.7%)
Site #2
79/93
89/93
91/93
92/93
1/93
(84.9%)
(95.7%)
(97.9%)
(98.9%)
(1.1%)
Site #3
1/1
1/1
1/1
1/1
0/1
(100%)
(100%)
(100%)
(100%)
(0.0%)
Site #4
19/25
24/25
25/25
25/25
0/25
(76.0%)
(96.0%)
(100%)
(100%)
(0.0%)