Table A-4 Venous Specimens
Analytical and System Accuracy Comparison: Glucose con-
centrations > 75 mg/dL
Site
Within
± 5%
Within
± 10%
Within
± 12.5%
Within
± 15%
Within
± 20%
Exceeds
± 20%
Site #1
77/112
108/112 111/112
111/112
112/112 0/112
(68.8%) (96.4%) (99.1%) (99.1%) (100%)
(0.0%)
Site #2
84/127
116/127 123/127 127/127 127/127 0/127
(66.1%) (91.3%) (96.9%) (100%)
(100%)
(0.0%)
Site #5
10/29
22/29
26/29
29/29
29/29
0/29
(34.5%) (75.9%) (89.7%) (100%)
(100%)
(0.0%)
Combined 171/268 246/268 260/268 267/268 268/268 0/268
All Sites (63.8%) (91.8%) (97.0%) (99.6%) (100%)
(0.0%)
Appendix