55
OK
Measured Power
2.1[mW]-3.5[mW]?
RDRV level
50[%] -150[%]?
RDRV level
150[%] over?
FPD level
50[%] under?
Check connection between conector
and FPC.
Check if FPC is not broken.
Yes
Execute ‘S4LPCHK.exe’ and choose ‘2.
Read Power Check’
Input value of the power meter by [µW]
For example: Measured power is 2.55
[µW] -> Input value is 2550.
Replace the P/U and execute LD
Power calibration.
Replace the FEP and execute LD
Power calibration.
Replace the P/U and execute
LD Power calibration.
Replace the FEP and execute
LD Power calibration
No
No
No
FPD level
150[%] over?
No
No
Yes
Yes
Yes
Yes
NG 1
NG 4
NG 3
NG 2
A