Troubleshooting
GA10218_002_C0 - 10/2016 - © Leybold GmbH
99
W46 Emission for cathode 2 can can
not be switched.
Signal MSIBE on MSV board is not active. Emission for cathode 2 can not be
switched on. PHOENIX L300i switches to cathode 1. Order a new ion
source.
Cathode 2 is defective.
MSV board is defective.
E47 Emission for both cathodes can
not be switched on!
Signal MSIBE on MSV board is not active. Emission can not be switched on.
Exchange the cathode by changing the ion source. After having exchanged
the ion source it must be possible to switch on both cathodes manually via
the service menu.
Replace ion source.
MSV board is defective.
E48 Anode heater is faulty!
Signal MSAFD on MSV board is active. Anode heater fuse has blown.
Replace fuse F2 on the MSV board.
E50 No communication with turbo
pump converter.
Clock from the frequency converter has failed. No communication to the fre-
quency converter.
Fuse F4 on the wiring backplane has blown.
Drive electronics Turbo Drive S is defective.
E51 Unknown TMP error
The frequency converter Turbo Drive S indicates an unknown error code.
Inform Leybold service
E52 TMP frequency is too low!
TMP frequency is too low!
Frequency converter is faulty.
Turbo molecular pump is faulty.
W53 Temperature at electronic unit is
too high (>59°C)
Ambient temperature too high.
Ventilation failure.
Air filter dirty and have to be changed.
E
54 Temperature at electronic unit is
too high (>60°C).
Ambient temperature is too high.
Internal ventilation has failed.
Air filters are dirty and must be exchanged.
W55 Temperature at electronic unit is
too low (< 2°C)
The temperature sensor on the wiring plane indicates T < 2 °C. Run-up time
for the fore vacuum pump will be longer.
Temperature sensor is faulty.
Ambient temperature is too low.
E56 Inlet pressure p1 too low!
Output voltage Pirani P1 U < 0,27 V.
Pirani sensor P1 is defective.
Pirani electronics on the I/O board is defective.
E58 fore vacuum pressure p2 too low! Output voltage Pirani P2 U < 0,27 V.
Pirani sensor P2 is defective.
Pirani electronics on the I/O board is defective.
No. Displayed Message
Description and possible solutions
Содержание 250000V01
Страница 104: ...Waste Disposal GA10218_002_C0 10 2016 Leybold GmbH 104...
Страница 105: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 10 7 Certificates Fig 43...
Страница 106: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 106...
Страница 108: ...Appendix 107 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 109: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum 108...
Страница 110: ...Appendix 109 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 111: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum110...