Operation
48
GA10218_002_C0 - 10/2016 - © Leybold GmbH
RS232
Main Menu > Settings > Interfaces > RS232
The calibration function of Leak Ware is not suitable with the use of a Phoenix.
Define PLC outputs
Main Menu > Settings > Interfaces > Define PLC outputs
The following relay outputs are available for further signal processing. The
maximum rating for the relay contacts is 60V AC/1A.
The contacts are numbered from bottom to top.
All pins of digital I/O, digital out and recorder must not be connected with
voltages higher than 60V DC/25V AC (to grounding equipment conductor) or
reach this threshold.
Description of the operation mode of the Digital Out. The pin assignment for
contacts 8 to 16 follows the same order as for pins 5 to 7.
The actual pin setting can be seen under Info / View internal data / page 7.
The following digital out signals are selectable.
Trigger 1; Trigger 2 and 3 analog Trigger 1
Is open in case Trigger Level 1 is exceeded or the machine is not in condition of
Softkey 2: Baudrate and end sign
Settings for the baudrate selectable between 1200 and 19200, as well as
endsign between CR+LF, CR or LF.
Softkey 3: Data, Parity, Stop bits
The settings for Data (7 or 8), Parity (Even, Odd, None) and Stop bits (1 or
2) can be selected
Softkey 7: RS 232 Protocol
The protocol from the RS 232 can be chosen between: L200 size,
Diagnostic, L200 Leak Ware and ASCII code.
Pin
Assignment
1
PLC in free selectable
2
PLC in free selectable
3
PLC in free selectable
4
GND
5 to 7
Digital out free selectable, 5 center contact, 6 normally open
contact, 7 normally closed contact
8 to 10
Digital out free selectable
11 to 13
Digital out free selectable
14 to 16
Digital out free selectable
RS232
Define PLC outputs
Caution
Содержание 250000V01
Страница 104: ...Waste Disposal GA10218_002_C0 10 2016 Leybold GmbH 104...
Страница 105: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 10 7 Certificates Fig 43...
Страница 106: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 106...
Страница 108: ...Appendix 107 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 109: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum 108...
Страница 110: ...Appendix 109 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 111: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum110...