![LEYBOLD 250000V01 Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/leybold/250000v01/250000v01_operating-manual_1890884012.webp)
Description
12
GA10218_002_C0 - 10/2016 - © Leybold GmbH
1.1.3 Vacuum Diagram PHOENIX L300i MODUL
The PHOENIX L300i MODUL has no roughing pump integrated as the other
models. Therefore it can be used with an external pump only. This pump can be oil
sealed or a dry version with a roughing capacity between 2.5 and 65 m
3
/h. This
pump has to be connected to the DN25 KF at the side or under the bottom of the
PHOENIX L300i MODUL.
Pos. Description
1
MS: Mass Spectrometer, Helium sensor (180° magnetic field mass
spectrometer)
2
Turbo molecular pump (TMP, provides high vacuum conditions in
the MS)
3
Pirani gauge P2 (fore vacuum pressure)
4
Diaphragm pump (provides the fore vacuum pressure for the TMP
and pumps down the parts under test)
5
Inlet Port
6
Pirani gauge P1 (inlet pressure)
V1 … V7: Electromagnetic Valves to control the gas flows
Fig. 4 Vacuum diagram PHOENIX L300i MODUL
MS
P
1
V
1
V
2a
V
3
V
2b
V
6
P
2
V
4a
external pump
V
5
V
7
1
2
4
5
3
6
Содержание 250000V01
Страница 104: ...Waste Disposal GA10218_002_C0 10 2016 Leybold GmbH 104...
Страница 105: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 10 7 Certificates Fig 43...
Страница 106: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 106...
Страница 108: ...Appendix 107 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 109: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum 108...
Страница 110: ...Appendix 109 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 111: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum110...