Troubleshooting
GA10218_002_C0 - 10/2016 - © Leybold GmbH
101
E68 TMP error: temperature conver-
ter too high (>75°C)
Maximum temperature for the converter has been exceeded.
Ambient temperature too high: Feed cooler air to the PHOENIX L300i
Fan defective: Replace the fan
Fore vacuum or high-vacuum pressure to high: reduce the inlet pressure of
the PHOENIX L300i
E69 TMP error: Run-up time out
Max. time after which the pump must enter its normal operation mode has
been exceeded.
Fore vacuum or high-vacuum pressure to high: Reduce the inlet pressure of
the PHOENIX L300i
Bearing defective: Inform Leybold service for repair
E70 TMP error: TMP motor tem-
perature too high (>90°C)
Max. motor temperature has been exceeded.
Fore vacuum or high-vacuum pressure to high: Reduce the inlet pressure of
the PHOENIX L300i
Fan defective: Replace the fan
Ambient temperature too high: Feed cooler air to the PHOENIX L300i.
Bearing defective: Inform Leybold service for repair
E71 TMP error: TMP could not be
identified
Pump could not be identified or no pump is connected.
Pump not connected to Turbo.Drive S: Check connecting cable
E72 Emission off (P1 too high)
Air inrush
E73 Emission off (p2 too high)
The emission is switched off as soon as the pressure P2 > 0.2 mbar or 1,5
mbar in the measurement mode. If after closing the inlet valve the pressure
drops again, the leak detector will revert to the standby mode.
Air inrush in the measurement mode.
W74 Error proportional valve or control
electronics
(PHOENIX L300i DRY only)
Proportional valve V4b or control electronics for proportional valve V4b
mounted on top-hat rail next to forepump non-existent or faulty.
W75 Maximum evacuation time
exceeded
Within the pre set evacuation time
the pressure of 100 mbar was not
reached.
"Test sample has got a GROSS leak." and "Wrong setting (too short) of max.
time of evacuation."
W76 Maximum evacuation time until
measurement exceededWithin
the pre set evacuation time mea-
surement mode was not reached.
"Test sample has got a GROSS leak." and "Wrong setting (too short) of max.
time of evacuation."
W77 Peak not in Range
The signal maximum has shifted to mass range alignment limits.
Signal of leak rate was instable during mass adjustment. Calibrate again.
Check the basic setting for the anode voltage through the service menu.
Check calibrated leak.
No. Displayed Message
Description and possible solutions
Содержание 250000V01
Страница 104: ...Waste Disposal GA10218_002_C0 10 2016 Leybold GmbH 104...
Страница 105: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 10 7 Certificates Fig 43...
Страница 106: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 106...
Страница 108: ...Appendix 107 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 109: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum 108...
Страница 110: ...Appendix 109 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 111: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum110...