Operation
70
GA10218_002_C0 - 10/2016 - © Leybold GmbH
Partial flow setup/pump configuration for the PHOENIX L300i MODUL:
Before setting up the parameters the partial flow mode has to be enabled in the
main menu under Settings > Vacuum settings > Vacuum ranges with softkey no. 6
partial flow enable and confirm with the soft key OK.
Forepump setup for PHOENIX L300i MODUL
Softkey 2: Pump setup
Setting for the forepump if the PHOENIX L300i MODUL
is operated with a
partial flow system.
Softkey 7: Partial flow setup
Options for setting up the partial flow system.
Softkey 2:
The entry of the nominal pumping speed of the partial flow pump can be
decreased. The minimum pumping speed is 4 m
3
/h.
Softkey 3: Forepump type
The PHOENIX L300i MODUL with partial flow system can be operated
with a dry forepump (for ex. Scroll pump) or a wet forepump (oil sealed).
This key is for choosing a dry (Scroll, piston) fore vacuum pump.
Softkey 5: ?
Help text
Softkey 6:
The entry of the nominal pumping speed of the partial flow pump can be
increased. The maximum pumping speed is 80m
3
/h.
Softkey 7: Forepump type
The PHOENIX L300i MODUL with partial flow system can be operated
with a dry (for ex. Scroll pump) forepump or a wet pump (oil sealed).
This key is for choosing a wet (oil sealed) fore vacuum pump.
Partial flow setup PHOENIX
L300i MODUL
Vacuum settings PHOENIX
L300i DRY MODUL
Содержание 250000V01
Страница 104: ...Waste Disposal GA10218_002_C0 10 2016 Leybold GmbH 104...
Страница 105: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 10 7 Certificates Fig 43...
Страница 106: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 106...
Страница 108: ...Appendix 107 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 109: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum 108...
Страница 110: ...Appendix 109 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 111: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum110...