Troubleshooting
100
GA10218_002_C0 - 10/2016 - © Leybold GmbH
E60 p2>10mbar after 5 minutes since
power on
Run-up time of the fore vacuum pump is too long.
Fore pump is faulty.
Valve V2 does not open.
Leak in the high vacuum system.
E61 Emission fail.
Emission could not be switched on. MSV subassembly indicates a fault.
MENB emission current not within range.
MSV board is defective.
Both cathodes are defective, replace ion source.
W62 Flow through capillary to low.
In the sniffer mode the intake pressure of the sniffer line is controlled. If the
pressure falls below the minimum limit, the flow through the capillary is too
low (contamination) or the capillary is blocked (foreign objects, particles).
The minimum limit can be set by the menu. Default value is 0.05 mbar.
Filter in the tip is clogged.
W63 Capillary broken
In the sniffer mode the intake pressure of the sniffer line is controlled. If the
pressure exceeds the maximum limit, the flow through the capillary is too
high (no leak tightness, broken capillary).
The maximum limit can be set by the menu. Default value is 0,15 mbar.
The capillary is broken or has been torn off.
E64 TMP error: Nominal speed has
been exceeded by over 10%
Nominal speed of the pump has been exceeded by over 10%
EMC problems: Check connecting cable, insert it properly. Switch the power
supply voltage off and then on again.
Turbo.Drive S faulty: Inform Leybold Service
E65 TMP error: Pass through time
exceeded
Max. time for passing through the critical frequencies has been exceeded.
Fore vacuum or high-vacuum pressure to high: Reduce the inlet pressure of
the PHOENIX L300i
Bearing defective: Inform Leybold service for repair
E66 TMP error: Bearing temperature
too high (>67°C)
Max. bearing temperature has been exceeded
Fore vacuum or high-vacuum pressure to high: Reduce the inlet pressure of
the PHOENIX L300i
Fan defective: Replace the fan
Ambient temperature too high: Feed cooler air to the PHOENIX L300i
Bearing defective: Inform Leybold service for repair
E67 TMP error: Short circuit in TMP-
motor or connecting cable
Short circuit in the pump’s motor or the connecting cable
Check to see if the connecting cable is undamaged, exchange if required.
Inform Leybold service in case of short circuit in TMP motor
No. Displayed Message
Description and possible solutions
Содержание 250000V01
Страница 104: ...Waste Disposal GA10218_002_C0 10 2016 Leybold GmbH 104...
Страница 105: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 10 7 Certificates Fig 43...
Страница 106: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 106...
Страница 108: ...Appendix 107 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 109: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum 108...
Страница 110: ...Appendix 109 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 111: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum110...