Pressure limits for sniff mode
Main menu > Settings > Monitoring functions > Pressure limits for sniff mode
This function is automatically activated in sniff mode. The pressure limits define an
upper and lower limit of the inlet pressure P1. If the pressure is not in this range
error messages are generated:
P > upper limit: Capillary broken
P < lower limit: Flow through capillary too low (Capillary blocked)
Softkey No. 6:
Increase change over threshold EVAC-GROSS
Selectable between 3 - 15 mbar
Softkey No. 7:
Increase change over threshold GROSS-FINE
Selectable between 0,05 - 0,2 mbar, respectively 0,05 - 0,1 mbar for
PHOENIX L300i DRY and PHOENIX L300i MODUL
Softkey No. 2:
Decreasing the maximum pressure, Upper limit is 0,15 mbar (default)
Softkey No. 3:
Decreasing the minimum pressure
Softkey No. 4:
Setting for Quicktest
Setting for use with the Quicktest, upper limit is 0,05 mbar, lower limit 0
mbar.
Pushing the button again for default setting.
Softkey No. 6
Increasing the maximum pressure
Softkey No. 7:
Increase the minimum pressure, Lower limit is 0,05 mbar (default)
Pressure limits for sniff
mode
Содержание 250000V01
Страница 104: ...Waste Disposal GA10218_002_C0 10 2016 Leybold GmbH 104...
Страница 105: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 10 7 Certificates Fig 43...
Страница 106: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 106...
Страница 108: ...Appendix 107 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 109: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum 108...
Страница 110: ...Appendix 109 GA10218_002_C0 10 2016 Leybold Vacuum...
Страница 111: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum110...