Maintenance
46
GA03108_002_C1 - 04/2018 - © Leybold
5.8 Maintenance Schedule
Process
Meas./test quantity Maintenance intervall
Remark
Check oil level
Min./max. oil level in
oil level glass
Before switching on and daily
Check oil level with the pump at standstill,
see Section 3.1.1
Check oil quality
Visual
Weekly
In the normal state LVO 100 and PFPE
are light, clear and transparent. LVO 210
is yellow, clear and transparent. In the
case of black oil an oil change is neces-
sary, see Section 5.2.
Check oil quality
Neutralisation value
(DIN 51 558)
For normal operating conditions annu-
ally
When pumping corrosive vapours, in
the case of much dust and cyclic ope-
ration weekly to quarter yearly
If the neutralisation value of LVO 100 is >
2, then an oil change will be required, see
Section 5.2.
Oil change
For normal operating conditions annu-
ally
When pumping corrosive vapours, in
the case of much dust and cyclic ope-
ration weekly to quarter yearly
See Section 5.2.
Check leak tight-
ness of the cooling
water connections
Quarter yearly
Clean motor fan
and cooling fins
The cleaning intervals will depend on
the ambient conditions.
See Section 5.3.