9. Troubleshooting
Problem
Brightfield
The specimen can not be brought into focus.
Darkfield
No definite DF contrast is possible.
The image is unevenly or not uniformly illuminated.
Undesirable stray light.
Cause/Remedy
Use the correct immersion medium.
Place the specimen on the stage with the
coverslip facing down.
Make sure that the cover glass thickness is
correct and that it suits the indication on the
objective.
Make sure that you are using an objective
with coverslip correction.
Adjust the correction ring on the objective if
present.
Be sure that a DF objective is being used.
The objective aperture is too high:
Maximum 0.7 for condenser S1
Maximum 0.4 for condenser S2328
If necessary, reduce the objective aperture
using the iris diaphragm on the objective.
Check the condenser centering.
The magnification is too weak.
Use a higher magnification.
Remove the condenser head or condenser
lenses.
Clean the specimen and neighboring lenses
(
→
p 95).