44
6. Assembly
6.12 Installation and Replacement of Incident
Light Lamps
Caution!
Light sources pose a potential irradiation
risk (glare, UV-radiation, IR-radiation).
Therefore, lamps have to be operated in
closed housings.
Make sure that the lamp housing has been
disconnected from the power supply. Unplug
the power plug and the power supply during
assembly.
During assembly work on xenon burners, al-
ways wear the supplied protective gloves
and face protection (Fig. 58) (risk of explosion).
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path
(blinding hazard).
Lamp Housing 106 z
This lamp housing is suitable for use with a 12V
100W halogen lamp or a variety of gas discharge
lamps.
Caution!
Make sure to follow the instructions and
safety notes of the lamp supplier.
Before changing lamps allow at least 30 min-
utes for cooling down!
Fig. 58
Protective gloves and mask
Fig. 59
Lamp housing 106 z L with Hg 100W lamp
1
Collector focusing
2
Vertical lamp adjustment
3
Horizontal lamp adjustment
4
Hg lamp mount
5
Reflector adjustment (not visible)
2
5
3
1
4