6. Assembly
Fig. 78
Rear view of stand
1
RS232 ports
2
2 x USB
3
4 x EXT.
4
XYZ control for SmartMove
5
Connection for Leica CTR6000 electronics box
6
Condenser cable
7
Lamp power cable
Fig. 77
Rear view of Leica CTR6000
1
AC power socket
2
XY Stage
socket for motorized stage
3
Direct interface
socket optional
4
Z Control
for separate focus control
5
XYZ Control
for SmartMove
6
Microscope
socket for microscope
7
12V, max 100W
for the lamp power cable of stand
8
DL: reset button
7
1
2
3
4
5
8
6
1
2
3
7
6
5
4
Note:
The CTR6000 electronics box must not be used
with other stands. The serial number of the associ-
ated stand has been recorded on the back of the
electronics box.
• Connect the
Microscope
(77.6) socket to the
back of the stand (78.5) using the 25-pin
microscope cable.
• Connect the SmartMove remote control module
to the
XYZ-Control
socket (77.5).
• Connect the motorized stage, if present, to the
XY-Stage
socket (77.2).
• Connect the lamp power cable (78.7) to the
12V,
max 100W
socket (77.7).
Caution!
Make sure that the plugs are correctly in-
serted and secured to prevent overheating
of the sockets.