48
6. Assembly
Fig. 66
Opening the fluorescence drawer
Fig. 64
Filter cube,
front side
Fig. 65
Filter cube,
back side
Fig. 67
Open fluorescence drawer
Fig. 68
Inserting or removing a filter cube
6.13 Equipping the Incident Light Turret Disk
The fluorescence drawer is located on the right
side of the stand. Before opening this drawer,
remove the cap below the drawer covering the
analyzer slot. Remove the analyzer if it is already
in the slot.
The replacement of individual cubes is more
convenient with the microscope switched on.
The position to be changed then automatically
turns to the outside and you can be sure that the
cube is positioned in the correct holder. You can
therefore postpone installing the filter cubes un-
til after the microscope has been switched on.
You can also insert the filter cubes while the in-
strument is switched off.
Press the white button next to the drawer. The
drawer will glide out into its initial position.
This is the only position in which the inner disk
for the cubes can be turned to a free or desired
position.
The positions in the turret disk are numbered.
Depending on your equipment, the individual fil-
ter and reflector cubes have already been as-
signed to specific positions at the factory. For
details, check the identification sheet included
with your order.