59
7. Start-up
Fig. 81
LeicaScreen
Initialization
7.2 Switching on the Microscope
• Switch on the power of the CTR6000
electronics box at the On/Off switch (80.1).
The signal lamp (80.2) is lit green when the
unit is ready. All motorized microscope
components will then run through an
initialization phase.
Note:
If a PC is connected, switch on
the electronics
box first
, and
then the computer
.
After the initialization (Fig. 81) is complete, the
LeicaScreen will display the microscope’s cur-
rent settings (Fig. 82).
If a component has not been installed correctly,
the LeicaScreen will display an error message.
See Troubleshooting chapter,
→
p. 91.
Components such as diaphragms, condensers,
light and phase contrast rings have been pre-
centered at the factory. It may be necessary to
correct the centering after the microscope has
been transported and assembled.
Before performing the required steps, please fa-
miliarize yourself with the LeicaScreen and the
controls.
Caution!
After turning on the gas discharge lamp, the
burner must be immediately adjusted. There-
fore,
do not
turn on the power supply unit
yet. First, work in transmitted light in order to
familiarize yourself with the microscope’s
controls.
Fig. 80
Front side
Leica CTR6000
1
On/Off switch
2
Signal lamp
1
2
Fig. 82
LeicaScreen
after
initialization