11. Trouble Shooting
Problem
Polarization
No polarization contrast is possible.
Fluorescence
The image is completely dark (no fluorescence).
The fluorescence is too weak.
Cause/Remedy
Bring the polarizer and analyzer into cross po-
sition until they reach maximum darkness
(without specimen) (
→
p. 74, 79).
Open the shutter (
→
p. 76).
Select the incident light axis (IL) (
→
p. 48).
Check the antigen-antibody combination.
Insert a new lamp (
→
p. 26f).
Insert the Booster Lens (
→
p. 36).
Center the lamp (
→
p. 54ff).
Insert a new lamp (
→
p. 26f).