7. Startup
Caution:
In older lamps, the structure of the arc is no
longer clearly recognizable. The image is
then more like that of a HG 50 lamp. The im-
age and mirror image can no longer be su-
perimposed exactly. In this case, align both
images.
• Using the collector, defocus the image with
the knob (53.6) until the arc image and mirror
image are no longer recognizable and the im-
age is homogeneously illuminated.
• Exchange the reflector cube for lamp adjust-
ment for the original filter cube.
Note:
Turn off the microscope before exchanging
the reflector cube.