33
6. Assembly
Note:
At the 4-fold and 5-fold filter turrets, the num-
bers are right below the mounting. At the 8-fold
turret, the numbers are on the bottom of the
disc.
• Push the retention pin (30a.2 or 30b.2) and
continue to turn the filter turret until you
reach the next locking position.
• Again make sure that the turret engages (re-
tention pin unlocks) and insert the next filter
and/or reflector cube as described above.
• When all filters and reflector cubes have been
inserted, close the front cover plate again.
Note:
At the Leica DM6000 M, 2 positions for bright
field and dark field reflector cubes may be set
(depending on configuration).
Note:
Filter cubes that are not included in shipment
have to be learned in with the Software Leica
Application Suite (LAS), module: Set-up.
6.8 Polarizer and Analyzer
ICT/P transmitted light polarizer
• Using the left clamping screw, fasten the ICT/P
transmitted light polarizer to the underside of
the condenser holder (Fig. 32).
• Make sure that the red index point on the front
of the polarizer is aligned with 0.
• If necessary, insert the compensators (
λ
- and
λ
/4 plates) into the polarizer’s receptacle
(Fig. 33).
Incident light polarizers:
R/P polarizer, rotating polarizer
L/ICR, R/ICR polarizer
• Remove the plug cap on the right side of the
incident light axis (Fig. 34).
• Insert the polarizer into the receptacle until it
latches in place.
Attention:
Insert the polarizer only in the front receptacle!
Fig. 32
Assembly of the ICT/P transmitted light polarizer
1
Clamping screw
!
1
Содержание DM5500 B
Страница 1: ...1 Leica DM5500B Leica DM6000B Leica DM6000M Operating Manual Bedienungsanleitung Manuel d utilisation...
Страница 3: ...3 Leica DM5500B Leica DM6000B Leica DM6000M Operating Manual...
Страница 92: ...3 Leica DM5500B Leica DM6000B Leica DM6000M Bedienungsanleitung...
Страница 181: ...3 Leica DM5500B Leica DM6000B Leica DM6000M Manuel d utilisation...