6. Assembly
Motorized polarizer
• A motorized polarizer is already installed and
ready for operation in the DIC condenser.
Transmitted light and incident light analyzer
• Remove the plug cap on the left side of the
stand.
• Insert the analyzer into the receptacle until it
latches in place (Fig. 35).
Motorized analyzer
• Insert the analyzer cube as described in sec-
tion 6.7 "Equipping the Incident Light Turret
Disc"
→
p. 32, in the corresponding position on
the filter turret. See the list provided (“Identifi-
cation Sheet”) for the correct position.
Fig. 34
Inserting the polarizer
1
The plug cap is replaced with the polarizer.
2
Knurled wheel for fine adjusting the DIC prisms
1
1
6.9 DIC Prisms
With the microscopes Leica DM5500 B and
DM6000 B the DIC objective prisms are already
mounted in the DIC turret above the objective
revolving nosepiece (Fig. 34.2).
2
Fig. 35
Inserting the analyzer
1
The plug cap is replaced with the analyzer.
Fig. 33
Inserting the compensators