![Leica DM5500 B Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/leica/dm5500-b/dm5500-b_operating-manual_1873295027.webp)
27
6. Assembly
Fig. 21
Lamp mount with 12 V 100 W halogen lamp
Inserting the 12V 100W halogen lamp into the
106 z lamp housing
• Unscrew the fastening screws of the cover
and lift up the cover (20.1).
• Unscrew the fastening screws of the lamp
mount (20.8) and pull out the mount (Fig. 21).
• Insert the lamp with the dust cover straight
into the socket until it stops.
Caution!
Do not remove the lamp’s dust cover until
you have installed the lamp. Avoid finger-
prints on the lamp.
• Remove the dust cover.
• Reinsert the lamp mount and retighten the
fastening screw (20.8).
• Close the lamp housing and retighten the fas-
tening screws.
• Place the lamp housing in the incident light
lamp housing receptacle (22.1) and fasten it
with the clamping screw on the side.
• Connect the lamp housing to the power supply
for incident light (symbol
) (22.4).
• Now connect the lamp power cable of the mi-
croscope (22.5) to the Leica CTR5500 or
CTR6000 electronics box (18.1, p. 25)
Fig. 22
Rear side of stand
1
Incident light lamp housing receptacle
2
Transmitted light lamp housing receptacle
3
12 V 100 W connection for transmitted light (symbol:
)
4
12 V 100 W connection for incident light (symbol:
)
5
Lamp power cable
1
2
3 4
5
Содержание DM5500 B
Страница 1: ...1 Leica DM5500B Leica DM6000B Leica DM6000M Operating Manual Bedienungsanleitung Manuel d utilisation...
Страница 3: ...3 Leica DM5500B Leica DM6000B Leica DM6000M Operating Manual...
Страница 92: ...3 Leica DM5500B Leica DM6000B Leica DM6000M Bedienungsanleitung...
Страница 181: ...3 Leica DM5500B Leica DM6000B Leica DM6000M Manuel d utilisation...