![LeCroy WavePro 7 Zi series Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/lecroy/wavepro-7-zi-series/wavepro-7-zi-series_getting-started-manual_1866367082.webp)
Getting Started Manual
WP7Zi-GSM-E RevB
82
Vertical:
Turn the
L
EVEL
knob in the TRIGGER control group to adjust
the trigger's vertical threshold.
Turn this knob to adjust the level of the trigger source or the
highlighted trace. Level defines the source voltage at which the
trigger will generate an event: a change in the input signal that
satisfies the trigger conditions.
Alternatively, in the Trigger dialog, you can touch inside the
Level
field and type in a value using the pop-up numeric keypad. Quickly
set a level of zero volts by touching the
Zero Level
button.
An arrow on the right side of the grid shows the threshold position.
This arrow is only visible if the trigger source is displayed.
Simple Triggers
Edge Trigger
E
DGE
T
RIGGER
S
ETUP
Setup your trigger by following the main steps explained in
Trigger Setup
(below) providing values for
Type
,
Setup
,
Level
fields, and using the
Optimize for High Frequency
checkbox as needed.
Trigger Setup
Various parts of the trigger dialog are used to setup your trigger. Many
fields on the dialog change based on your Trigger Type selections.
S
ETUP