Getting Started Manual
WP7Zi-GSM-E RevA
170
Severe cases may require a reloading of the base operating system and
oscilloscope application. LeCroy provides you with an Acronis recovery
application and a backup image in an extra partition on the instrument’s
hard drive.
PLEASE NOTE THE FOLLOWING
:
The recovery process produces a replica of the operating system
and oscilloscope application software at the current revision levels
when the oscilloscope was manufactured. Any further revisions of
the application software, Windows operating system, and virus scan
definition files are not automatically upgraded. Therefore, after
completion of the disk image recovery, it is highly recommended to
search vendor Web sites and upgrade the individual components to
current revision levels.
After the recovery procedure is completed, you must activate
Windows, either by Internet connection to Microsoft’s Web site or
by telephone. Have your Windows Product Key number (located on
the rear of the oscilloscope) handy during Widows reactivation.
The current oscilloscope application software can be downloaded
directly from the LeCroy website at
www.lecroy.com
. Since the
calibration data for the oscilloscope is stored in the D: drive, current
calibration constants are not overwritten during the recovery
process.