WavePro 7 Zi and Zi-A Oscilloscopes
77
WP7Zi-GSM-E RevB
SMART Triggers
– sophisticated triggers that enable you to use
basic or complex conditions for triggering. Use SMART Triggers for
signals with rare features, like glitches.
Cascade Trigger
- a combination of a sequence of 2 to 4 trigger
stages to establish precise cause and effect trigger conditions.
Trigger Types
There are a set of standard Trigger Types. These Triggers are basic
waveform features or conditions such as a positive or negative slope, and
hold-off and are available on all oscilloscopes along with tools like
Software Assisted Trigger
and
Trigger Scan
.
In addition, there are more sophisticated triggers for setting basic or
complex conditions. These include
Smart Triggers
,
Serial Trigger
,
Qualified
Triggers
, and
Cascade Triggers
.
Learn more and
Contact LeCroy for Support
(on page 177).
Edge
A simple trigger, Edge trigger is activated by basic waveform features or
conditions such as positive or negative slope, and holdoff.
Width
Width
trigger allows you to define a positive- or negative-going pulse
width bounded by a voltage level, above or below which a trigger will
occur. Or you can specify a pulse width and voltage range, within or
outside of which a trigger will occur.
Qualified A-B
In single trigger mode,
Qualify A-B
arms the oscilloscope on the A event,
and then triggers on the B event.