WavePro 7 Zi and Zi-A Oscilloscopes
153
WP7Zi-GSM-E RevB
You can log all events or errors only. This log can be invaluable when you
are creating and debugging remote control applications.
Remote Communication Setup
If you are connecting the oscilloscope to a network, first contact your
Information Systems administrator. If you are connecting the oscilloscope
directly to your PC, connect a GPIB or Ethernet cable between them.
1.
Touch
Utilities → Utilities Setup
on the menu bar.
2.
Touch the
Remote
tab.
3.
Make a
Port
selection:
TCPIP
(transmission control
protocol/Internet protocol) or
GPIB
(general purpose interface bus).
If you do not have a GPIB card installed, the GPIB selection will not
be accessible.
4.
If you are using GPIB, set a GPIB address by touching inside the
GPIB
Address
data entry field and enter an address.
OR
If you are using TCPIP, press the
Net Connections
button; the
Windows
Network Connections
window opens. You can reconfigure
the oscilloscope's connection if it is already connected to the
network or make a new connection.
Note
: Your instrument allows you to restrict remote control access to
certain clients. To restrict access, under
Security
, touch the
Yes
button
and enter the IP addresses or domain name server names you want to
restrict separated by a comma.
Configuring the Remote Control Assistant Event Log
1.
Touch
Utilities → Utilities Setup
on the menu bar.
2.
Touch the
Remote
tab.
3.
Under
Remote Control Assistant
, touch inside the
Log Mode
data
entry field. Select
Off
,
Errors Only
, or
Full Dialog
from the pop-up
menu.