Installation 2-43
instrument resides. To complete the chassis and slot assignment,
click on OK. The dialog box will be removed from the screen the and
Slot configuration parameters will be updated with the information you
programmed just now. Notice the 10-digit number above the slot
number. This is the serial number of your instruments as was
detected by the ArbDetector. To make sure this is the right instrument
you want to allocate, remove the instrument from your chassis and
compare to the serial number as printed on the card.
Testing
Communications
With your
Instrument
Final operation with your ArbDetector allows you to test
communication with your instrument. If you click on the
Communication… tab, the tab, as shown in Figure 2-30 will show. The
Command Editor field lets you write SCPI commands that control
instrument functions and parameters. SCPI programming is covered
in Chapter 4 of this manual. Connect your instrument to an
oscilloscope and try these simple commands just to see you’re your
instrument responds correctly:
output on
will turn on your output and sinewave will be
frequency 2e6
will change output frequency to 2 MHz
volt 5
will change output amplitude to 5 Vp-p
If you are not sure about the commands and want the Command
Editor to alert on errors, check the Add “Syst:Err?” Query option. If
you make syntax errors, the errors will display in the Response field.
The example in Figure 2-30 shows 0, “No error”. Different digits and
error description will be the response if you make errors.
The Response field can also be used to query instrument settings.
Try the following queries and check the response:
output?
will query output state
frequency?
will query frequency setting
volt?
will query amplitude setting
The commands as you type them will be saved as an history list. You
can use any command from this list by unfolding the list and selecting
a command, then clicking on Send to program this command.
Содержание PXA125
Страница 14: ...User Manual PXA125 xiv This page intentionally left blank ...
Страница 58: ...User Manual PXA125 2 44 Installation Figure 2 30 Communicating with your Instrument ...
Страница 118: ...User Manual PXA125 3 104 ArbConnection This page intentionally left blank ...
Страница 120: ...User Manual PXA125 Programming Reference 4 2 ...
Страница 161: ...User Manual PXA125 Programming Reference 4 43 ...
Страница 167: ...User Manual PXA125 Programming Reference 4 49 MAXimum 10 in units of volts ...
Страница 179: ...User Manual PXA125 Programming Reference 4 61 ...
Страница 186: ...User Manual PXA125 Programming Reference 4 68 ...
Страница 193: ...User Manual PXA125 Programming Reference 4 75 ...
Страница 208: ...User Manual PXA125 Programming Reference 4 90 This page intentionally left blank ...
Страница 219: ...User Manual PXA125 Specifications A 101 This page intentionally left blank ...