User Manual PXA125
Programming Reference 4-81
The Status Byte
Register (STB)
The Status Byte summary register contains conditions from the other
registers. Query data waiting in the generator's output buffer is
immediately reported through the Message Available bit (bit 4). Bits in
the summary register are not latched. Clearing an event register will
clear the corresponding bits in the Status Byte summary register.
Description of the various bits within the Status Byte summary register
is given in the following:
Bit 0
- Decimal value 1. Not used, always set to 0.
Bit 1
- Decimal value 2. Not used, always set to 0.
Bit 2
- Decimal value 4. Not used, always set to 0.
Bit 3
- Decimal value 8. Not used, always set to 0.
Bit 4
- Decimal value 16. Message Available Queue Summary
Message (MAV). The state of this bit indicates whether or not the
output queue is empty. The MAV summary message is true when
the output queue is not empty. This message is used to
synchronize information exchange with the controller. The
controller can, for example, send a query command to the device
and then wait for MAV to become true. If an application program
begins a read operation of the output queue without first checking
for MAV, all system bus activity is held up until the device
responds.
Bit 5
- Decimal value 32. Standard Event Status Bit (ESB)
Summary Message. This bit indicates whether or not one or more
of the enabled ESB events have occurred since the last reading
or clearing of the Standard Event Status Register.
Bit 6
- Decimal value 64. Master Summary Status
(MSS)/Request Service (RQS) Bit. This bit indicates if the device
has at least one condition to request service. The MSS bit is not
part of the IEEE-STD-488.1 status byte and will not be sent in
response to a serial poll. However, the RQS bit, if set, will be sent
in response to a serial poll.
Bit 7 - Decimal value 128. Not used, always set to 0.
Содержание PXA125
Страница 14: ...User Manual PXA125 xiv This page intentionally left blank ...
Страница 58: ...User Manual PXA125 2 44 Installation Figure 2 30 Communicating with your Instrument ...
Страница 118: ...User Manual PXA125 3 104 ArbConnection This page intentionally left blank ...
Страница 120: ...User Manual PXA125 Programming Reference 4 2 ...
Страница 161: ...User Manual PXA125 Programming Reference 4 43 ...
Страница 167: ...User Manual PXA125 Programming Reference 4 49 MAXimum 10 in units of volts ...
Страница 179: ...User Manual PXA125 Programming Reference 4 61 ...
Страница 186: ...User Manual PXA125 Programming Reference 4 68 ...
Страница 193: ...User Manual PXA125 Programming Reference 4 75 ...
Страница 208: ...User Manual PXA125 Programming Reference 4 90 This page intentionally left blank ...
Страница 219: ...User Manual PXA125 Specifications A 101 This page intentionally left blank ...