User Manual PXA125
Programming Reference 4-88
received, but the instrument does not accept one in this position.
-131,"Invalid suffix". A suffix was incorrectly specified for a numeric
parameter. The suffix may have been misspelled.
-148,"Character data not allowed". A character data element was
encountered where prohibited by the instrument.
-200,"Execution error". This is the generic syntax error for the
instrument when it cannot detect more specific errors. Execution error
as defined in IEEE-488.2 has occurred.
-221,"Setting conflict". Two conflicting parameters were received
which cannot be executed without generating an error. Listed below
are events causing setting conflicts.
1. Sum of pulse or ramp parameters is more than 100. Corrective
action: Change parameters to correct the problem.
2. ampl/2 + |offset| is more than 4. Corrective action: Reduce offset
to 0, then change amplitude-offset values to correct the problem.
3. Activating filters when the PXA125 is set to output the built-in sine
waveform, or activating the built-in sine waveform when one of
the PXA125 filters is turned on. Corrective action: If in sine, select
another function and activate the filter(s).
4. Activating burst mode when the PXA125 is set to sequence
mode, or activating sequence mode when the PXA125 is set to
burst mode. Corrective action: Remove the PXA125 from burst or
sequence and then selected the desired mode.
•
Changing operating mode from triggered to continuous when the
PXA125 is set to single sequence advance, or changing the
operating mode from continuous to triggered when the PXA125
is set to automatic sequence advance mode. Corrective action:
Observe the PXA125 advance mode while setting sequence
advance.
•
Turning on FM, FSK, SWP, or Arm when one function is already
on.
•
Selecting TTLTrig line as a source advance when it has already
been selected as an output.
-222,”Data out of range”. Parameter data, which followed a specific
header, could not be used because its value is outside the valid range
defined by the generator.
-224,”Illegal parameter value”. A discrete parameter was received
which was not a valid choice for the command. An invalid parameter
choice may have been used.
Содержание PXA125
Страница 14: ...User Manual PXA125 xiv This page intentionally left blank ...
Страница 58: ...User Manual PXA125 2 44 Installation Figure 2 30 Communicating with your Instrument ...
Страница 118: ...User Manual PXA125 3 104 ArbConnection This page intentionally left blank ...
Страница 120: ...User Manual PXA125 Programming Reference 4 2 ...
Страница 161: ...User Manual PXA125 Programming Reference 4 43 ...
Страница 167: ...User Manual PXA125 Programming Reference 4 49 MAXimum 10 in units of volts ...
Страница 179: ...User Manual PXA125 Programming Reference 4 61 ...
Страница 186: ...User Manual PXA125 Programming Reference 4 68 ...
Страница 193: ...User Manual PXA125 Programming Reference 4 75 ...
Страница 208: ...User Manual PXA125 Programming Reference 4 90 This page intentionally left blank ...
Страница 219: ...User Manual PXA125 Specifications A 101 This page intentionally left blank ...