User Manual PXA125
Specifications A-91
SPECIFICATIONS
Output Channels
1
Multiple Instrument synchronization
Description
Multiple instruments can be connected together and synchronized to
provide multi-channel synchronization.
Sample Clock
Source From Master card to slave boards through the local bus
Range and Resolution Same as Sample Clock but limited to 80 Ms/s
Initial Skew
<20 ns to the first master; 20 ns cumulative to additional
slaves
Phase Offset Between Instruments
Programmable from 0 to n points
SAMPLE CLOCK
Range 100 mS/s to 125 MS/s
Resolution
10 digits limited by 1
µ
Hz
Accuracy
Same as reference
StabilitySame as reference
Reference
Internal System clock,
≥
0.01% (100 ppm) initial tolerance over a 19
°
C to
29
°
C temperature range; 100ppm/
°
C below 19
°
C and above 29
°
C; 100ppm/year aging rate
TCXO
≥
0.0001% (1 ppm) initial tolerance over a 19
°
C to 29
°
C temperature
Содержание PXA125
Страница 14: ...User Manual PXA125 xiv This page intentionally left blank ...
Страница 58: ...User Manual PXA125 2 44 Installation Figure 2 30 Communicating with your Instrument ...
Страница 118: ...User Manual PXA125 3 104 ArbConnection This page intentionally left blank ...
Страница 120: ...User Manual PXA125 Programming Reference 4 2 ...
Страница 161: ...User Manual PXA125 Programming Reference 4 43 ...
Страница 167: ...User Manual PXA125 Programming Reference 4 49 MAXimum 10 in units of volts ...
Страница 179: ...User Manual PXA125 Programming Reference 4 61 ...
Страница 186: ...User Manual PXA125 Programming Reference 4 68 ...
Страница 193: ...User Manual PXA125 Programming Reference 4 75 ...
Страница 208: ...User Manual PXA125 Programming Reference 4 90 This page intentionally left blank ...
Страница 219: ...User Manual PXA125 Specifications A 101 This page intentionally left blank ...