![LeCroy DDA-3000A Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/lecroy/dda-3000a/dda-3000a_operators-manual_1866311054.webp)
52
WM-OM-E Rev I
Trigger Knobs:
Level
Selects the trigger threshold level. The Level is indicated in the
Trigger
label:
Trigger Buttons:
Setup
Activates the trigger setup menu to select the trigger type and the trigger
conditions.
Stop
Prevents the scope from triggering on a signal. If you boot up the
instrument with the trigger in Stop mode, the message "no trace available"
will be displayed. Press Auto to display your trace.
Auto
Triggers the scope after a time-out, even if the trigger conditions are not
met.
Normal
Triggers the scope each time a signal is present that meets the conditions
set for the type of trigger selected.
Single
Arms the scope to trigger once (single-shot acquisition) when the input
signal meets the trigger conditions set for the type of trigger selected. If
the scope is already armed, it will force a trigger.
Horizontal Knobs:
Delay
Horizontally positions the scope trace on the display so you can observe
the signal prior to the trigger time.
Delay
adjusts the pre- and post-trigger
time.
Time/Division
Sets the time/division of the scope timebase (acquisition system).
LeCroy's SMART Memory feature automatically optimizes the memory
and sample rate for maximum resolution.
Horizontal Buttons:
Smart Memory
Calls up the SMART Memory dialog from the Horizontal setup menus.
Zero Delay
Sets the horizontal delay to zero. The trigger point is positioned in the
middle of the display grid.
Setup
Activates the TIMEBASE menu to allow you to select acquisition
conditions, including the sample mode, maximum memory length, etc.
Vertical Knobs:
Offset
Adjusts the vertical offset of a channel.
Содержание DDA-3000A
Страница 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Страница 18: ...16 WM OM E Rev I BLANK PAGE ...
Страница 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Страница 65: ...X Stream Operator s Manual WM OM E Rev I 63 then Math Setup then Measure Setup then Utilities Setup ...
Страница 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Страница 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Страница 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Страница 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Страница 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Страница 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Страница 284: ...282 WM OM E Rev I ...
Страница 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Страница 321: ...X Stream Operator s Manual WM OM E Rev I 319 Further Examples of MATLAB Waveform Functions Negate the input signal ...
Страница 324: ...322 WM OM E Rev I ...