X-Stream Operator’s Manual
WM-OM-E Rev I
47
Installation (Overvoltage) Category II refers to local
distribution level, which is applicable to equipment
connected to the mains supply (AC power source).
Installation (Overvoltage) Category I refers to signal
level, which is applicable to equipment measuring
terminals that are connected to source circuits in which
measures are taken to limit transient voltages to an
appropriately low level.
Pollution Degree 2 refers to an operating environment
where normally only dry non-conductive pollution
occurs. Occasionally a temporary conductivity caused
by condensation must be expected.
Protection Class 1 refers to a grounded equipment, in
which protection against electric shock is achieved by
Basic Insulation and by means of a connection to the
protective ground conductor in the building wiring.
Note
The design of the instrument has been verified to
conform to
EN 61010-1 safety standard per the following
limits:
Installation (Overvoltage) Categories II (Mains
Supply Connector) & I (Measuring Terminals)
Pollution Degree 2
Protection Class I
Cooling
The instrument relies on forced air cooling with internal
fans and ventilation openings. Care must be taken to
avoid restricting the airflow around the apertures (fan
holes) at the sides and rear of the scope. To ensure
adequate ventilation it is required to leave a 10 cm (4
inch) minimum gap around the sides and rear of the
instrument.
CAUTION
Do not block the ventilation holes
located on both sides and rear of the
scope.
The instrument also has internal fan control circuitry that
regulates the fan speed based on the ambient
temperature. This is performed automatically after
start-up with no manual intervention required.
CAUTION
Do not allow any foreign matter to
enter the scope through the
ventilation holes, etc.
AC Power Source
100 to 240 Vrms (+/-10%) AC at 50/60 Hz; 115 Vrms
(+/-10%) AC at 400 Hz; Automatic AC voltage selection;
Installation Category: 300V CAT II
No manual voltage selection is required because the
instrument automatically adapts to line voltage.
WaveMaster 8620A, 8420
</= 800 watts (800 VA)
depending on accessories
installed (internal printer,
probes, PC port plug-ins,
etc.)
Note:
The instrument automatically adapts itself to the
AC line input within the following ranges:
Voltage
Range
90 to 132 VAC 180 to 264
VAC
Frequency
Range
45 to 440 Hz
45 to 66 Hz
Содержание DDA-3000A
Страница 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Страница 18: ...16 WM OM E Rev I BLANK PAGE ...
Страница 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Страница 65: ...X Stream Operator s Manual WM OM E Rev I 63 then Math Setup then Measure Setup then Utilities Setup ...
Страница 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Страница 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Страница 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Страница 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Страница 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Страница 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Страница 284: ...282 WM OM E Rev I ...
Страница 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Страница 321: ...X Stream Operator s Manual WM OM E Rev I 319 Further Examples of MATLAB Waveform Functions Negate the input signal ...
Страница 324: ...322 WM OM E Rev I ...