![LeCroy DDA-3000A Скачать руководство пользователя страница 210](http://html1.mh-extra.com/html/lecroy/dda-3000a/dda-3000a_operators-manual_1866311210.webp)
208
WM-OM-E Rev I
DDA-5005A.
TAA-
Average local baseline-to-trough
amplitude for all local features.
Hysteresis argument
used to discriminate
levels from noise in
data.
Available with DDM2
option.
Standard in
DDA-5005A.
TAA+
Average local baseline-to-peak
amplitude for all local features.
Hysteresis argument
used to discriminate
levels from noise in
data.
Available with DDM2
option.
Standard in
DDA-5005A.
TIE@level
Difference between the measured
times of crossing a given slope and
level and the ideal expected time.
For Slope you can choose positive,
negative, or both. For output units
you can choose time or unit interval
(UI). A unit interval equals one clock
period.
The Virtual Clock setup gives you a
choice of Standard (1.544 MHz) or
Custom reference clocks.
You can also use a mathematically
derived Golden PLL to filter low
frequency jitter. The cutoff
frequency is user selectable.
Cutoff Freq = (1/1.667e3) x
Clock Freq
Reference levels and
edge-transition polarity
can be selected.
Hysteresis argument
used to discriminate
levels from noise in
data.
Available with JTA2,
ENET, SDA, and XMAP
options.
Standard in SDA100G,
WavePro 7000A,
WaveExpert, and
sampling scopes.
Time@edge
Measures time at each edge on
each digital line
Available with MS-32
option.
Time@level
Time at level: Time from trigger
(t=0) to crossing at a specified level.
Time from trigger to crossing
level
Reference levels and
edge-transition polarity
can be selected.
Hysteresis argument
used to discriminate
levels from noise in
data.
Standard parameter.
Top
Higher of two most probable states,
the lower being base; it is
Value of most probable
higher state
Gives similar result
when applied to time
Содержание DDA-3000A
Страница 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Страница 18: ...16 WM OM E Rev I BLANK PAGE ...
Страница 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Страница 65: ...X Stream Operator s Manual WM OM E Rev I 63 then Math Setup then Measure Setup then Utilities Setup ...
Страница 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Страница 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Страница 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Страница 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Страница 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Страница 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Страница 284: ...282 WM OM E Rev I ...
Страница 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Страница 321: ...X Stream Operator s Manual WM OM E Rev I 319 Further Examples of MATLAB Waveform Functions Negate the input signal ...
Страница 324: ...322 WM OM E Rev I ...