8. Under "Buffer," touch inside the
#Values
data entry field and enter a value from 20 to 1000.
9. Under "Scaling," touch inside the
#Bins
data entry field and enter a value from 20 to 2000.
10. Touch the
Find Center and Width
button to center the histogram. Or touch inside the
Center
, then the
Width
, data entry fields and enter a value using the pop-up numeric
keypad.
11. Touch inside the Vertical Scale field and select
Linear
or
Linear Constant Max
from the
pop-up menu:
.
To View Thumbnail Histograms
Histicons are miniature histograms of parameter measurements that appear below the grid. These
thumbnail histograms let you see at a glance the statistical distribution of each parameter.
1. In the menu bar, touch
Measure
, then one of the Measure Mode buttons:
Std Vertical
,
Std
Horizontal
, or
My Measure
.
2. Touch
the
Histicons
checkbox to display thumbnail histograms below the selected
parameters.
Note:
For measurements set up in My Measure, you can quickly display an enlarged histogram of a thumbnail histogram by
touching the Histicon you want to enlarge. The enlarged histogram will appear superimposed on the trace it describes. This
does not apply to "Std Vertical" or "Std Horizontal" measurements.
Persistence Histogram
You can create a histogram of a persistence display also by cutting a horizontal or vertical slice
through the waveform. You also decide the width of the slice and its horizontal or vertical placement
on the waveform.
This math operation is different than the "Histogram" math operation and is not affected by
Center
and
Width
settings made there.
To Set Up Persistence Histograms
1. In the menu bar, touch
Math
, then
Math Setup
.
2. Touch one of function tabs
F1
through
Fx
The number of math traces available depends on
the software options loaded on your scope. See specifications..
3. Touch
inside
the
Source1
field and select a source from the pop-up menu.
4. Touch
inside
the
Operator1
field and select
Phistogram
from the
Select Math
Operator
menu.
Содержание DDA-3000A
Страница 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Страница 18: ...16 WM OM E Rev I BLANK PAGE ...
Страница 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Страница 65: ...X Stream Operator s Manual WM OM E Rev I 63 then Math Setup then Measure Setup then Utilities Setup ...
Страница 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Страница 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Страница 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Страница 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Страница 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Страница 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Страница 284: ...282 WM OM E Rev I ...
Страница 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Страница 321: ...X Stream Operator s Manual WM OM E Rev I 319 Further Examples of MATLAB Waveform Functions Negate the input signal ...
Страница 324: ...322 WM OM E Rev I ...