- 18 -
While the net weight is displayed, keep
pressed to display the preset tare. When
the key is released the net weight will be displayed again.
-
IF A SEMI-AUTOMATIC TARE (NET) IS ENTERED, IT IS NOT POSSIBLE TO
ACCESS THE ENTER PRESET TARE FUNCTION.
-
IF A PRESET TARE IS ENTERED, IT’S STILL POSSIBLE TO ACCESS
THE SEMI-
AUTOMATIC TARE (NET) FUNCTION. THE TWO DIFFERENT TYPES OF TARE ARE
ADDED.
ALL THE SEMI-AUTOMATIC TARE (NET) AND PRESET TARE FUNCTIONS WILL BE
LOST WHEN THE INSTRUMENT IS TURNED OFF.
Function not available if net functions are disabled (see parameter
in section
INFO MENU
).
SEMI-AUTOMATIC ZERO (WEIGHT ZERO-SETTING FOR SMALL VARIATIONS)
By closing the SEMI-AUTOMATIC ZERO input, the weight is set to zero; alternatively, by pressing
the
key for less than 3 seconds, the
message is displayed for 3 seconds, by
pressing ENTER the weight is set to zero. The zero-setting will be lost upon instrument power-off.
This function is only allowed when:
-
the weight is stable; otherwise the
alarm appears;
-
the combined effect of the semi-automatic zero and zero tracking does not exceed 2% of the
maximum capacity, otherwise the alarm
appears.
0