- 56 -
READING OF VALUES STORED IN ALIBI MEMORY
To recall a stored value from the alibi memory:
-
write the identification number of the value to recall in the “Alibi memory identification number”
register;
-
send the command 111 to the Command Register;
-
read the data from the 40084 ÷ 40090 registers:
-
“Weight read from the alibi memory” register
(40084 - 40085): gross weight or net weight
(check the “Net weight” bit in the table
TYPE OF DATA READ FROM THE ALIBI MEMORY
REGISTER
to determine whether it is net or gross);
-
“Tare read from the alibi memory” register
(40086 - 40087): when the value is equal to zero,
it means that you are reading a gross weight, otherwise you are reading a net weight;
-
“Decimals read from the alibi memory” register
(40088): number of decimals to apply to
weight values;
-
“Unit of measure read from the alibi memory” register
(40089): unit of measure code (see
table in
DIVISIONS AND UNITS OF MEASURE REGISTER
section for the codes legend);
-
see the table
TYPE OF DATA READ FROM THE ALIBI MEMORY REGISTER
(40090) for
a description of the same-named register; use this register to check if the weight reading
refers to a net weight and if the tare reading is a preset tare;
-
if the requested value does not exist, all the registers from 40084 to 40090 will be set to zero.
TYPE OF DATA READ FROM THE ALIBI MEMORY REGISTER (40090)
Bit 0
The read weight is a net weight
Bit 8
Bit 1
The read tare is a preset tare
Bit 9
Bit 2
Bit 10
Bit 3
Bit 11
Bit 4
Bit 12
Bit 5
Bit 13
Bit 6
Bit 14
Bit 7
Bit 15