LAUDA Ultra-Thermostats
UB 20(-D), UB 25, UB 50, UB 30, UB 40
UB 20 F, UB 20 J, UB 30 J, UB 40 J, UB 65 J
UB 20 JL, UB 30 JL, UB 40 JL
YATE0013 /10.04.02
- 8 -
Technical data to DIN 58966
UB 20 F
UB 20 J
UB 30 J
UB 40 J
UB 65 J
Operating temperature range
(°C)
-30...200
-30...300
-30...300
-30...300
-30...300
(with external cooling)
Working temperature range
Simplex
(°C)
35...200
45...300
45..300
45...300
40...300
with water cooling
(°C)
20...200
20...300
20...300
20...300
20...300
Temperature sensor
Pt 100 Class B to DIN IEC 751
Temperature variation at
70°C in the bath
(°C)
±
0.005*)
±
0.005...
±
0.01*)
Heater power (max.)
(kW)
1,2
3
3
3
3
Safety system
2nd resistance thermometer and level sensor
Simplex pump
Pump output against
zero head
(outside the bath)
(l/min)
12
15
15
15
15
Pump pressure (max.)
(bar)
0.2
0.3
0.3
0.3
0.3
Pump connections
M 16 x 1; nipples 13
∅
Filling volume
(l)
15...18
15...18
22.5...30.5
32...40.5
48...54
Bath opening
(mm)
∅
195
∅
215
Bath depth
(mm)
195
195
320
450
690
Usable liquid depth
(mm)
175
175
300
430
650
Height to top of bath
(mm)
265
265
390
520
755
Bench area (W x D) x Height
(mm)
(300x450)x465 (300x450)x465 (300x450)x590
(300x450)x720 (320x485)x955
(only thermostat)
Weight (with R 400-P)
(kg)
27
27
33
39
60
Supply
230 V; 50 Hz / 230 V; 60 Hz
(kW)
1.4
3.2
3.2
3.2
3.3
Protection Class 1 to VDE 0106
Units are conform to EU Guideline 89/336/EWG (EMC) and 73/23/EWG (low-voltage) and carry the CE mark (230 V; 50 Hz).
Cat. No. compl. with R 400-P
230 V; 50 Hz
LTB 139
LTB 136
LTB 137
LTB 138
LTB 142
230 V; 60 Hz
LTB 239
LTB 236
LTB 237
LTB 238
LTB 242
*) see Item 4.3