LAUDA Ultra-Thermostats
UB 20(-D), UB 25, UB 50, UB 30, UB 40
UB 20 F, UB 20 J, UB 30 J, UB 40 J, UB 65 J
UB 20 JL, UB 30 JL, UB 40 JL
YATE0013 /10.04.02
- 44 -
The display shows
L
1
PROG. INP
L
2
NO OF CYCLES:_
Input 1 ... 99 is possible.
With more than one cycle it is convenient to have the final temperature and the starting
temperature T
start
at the same level!
Afterwards a tolerance range can be input for monitoring the programme.
The display shows
L
1
PROG. INP
L
2
TOL. RANGE_.
Now you can input a tolerance range value from 0.1 to 9.9°C. I.e. if the control variable
(bath temperature or external temperature T
1
or T
2
) deviates from the set temperature
of the segment by more than the tolerance range value while the programme is running,
the proramme sequence will be stopped until the control variable is within the tolerance
range again.
At the same time a "T" appears on the right in L
2
.
The input of 0.0 switches off the tolerance range function.