LAUDA Ultra-Thermostats
UB 20(-D), UB 25, UB 50, UB 30, UB 40
UB 20 F, UB 20 J, UB 30 J, UB 40 J, UB 65 J
UB 20 JL, UB 30 JL, UB 40 JL
YATE0013 /10.04.02
- 6 -
When connecting up an external system, check that filling this system does not cause
the level inside the thermostat to fall more than is permitted.
When the thermostating liquid has reached the setpoint the symbol
starts to flash on
the right in line 1 (L
1
) of the display.
After the unit has stabilised the bath temperature (T
i
) corresponds to the setpoint (T
S
).
1.2 Operating
safety
The thermostat must be operated with non-flammable bath liquids only, or with
flammable bath liquids up to 25°C below their fire point, otherwise there is a possibility
that a flammable atmosphere may form (see Item 4.2).
1.3 WARNING:
Parts of the bath cover may reach temperatures above 70°C when working at higher
temperatures.The outflow and return pipes of the pumps reach the operating
temperature. Touching them is dangerous because of very high or low temperatures.