Retrofit kit for dial indicator test devices
ed. 01/2017
page 11
Now carefully position the dial indicator test device in the designated area of the base plate („between“
the stops).
ATTENTION:
Observe the relevant safety regulations when lifting the dial indicator test device! Avoid any collision with
the lighting unit and the camera!
Position the dial indicator test device on the base plate, making sure that the base of the test instrument
fits well in between the stops on the front and on the right side of the base plate. Now replace the
unfastened stops and fix the position of the dial indicator test device by fitting the stops tightly to the
socket of the test device and re-fasten the fixing screws.
Fig. 6: front stops
To position the lighting unit (see Fig. 7), loosen the clamp (1) of the horizontal slide of the lighting unit (2)
and move the slide horizontally until the rear side of the lighting unit is positioned in a distance of 1 – 2
mm from the camera lens (see Fig. 8). Fix the slide in this position by tightening the clamp (1).