Model 4500 Premier Dust and Opacity Monitor
79
8.
Lock the sliding adjusting plate in place.
IMPORTANT
The M203 Clear Stack Simulator once calibrated to an instrument should not be adjusted
unless undergoing another re-calibration.
9.
Undo the 2 socket head screws and remove the Simulator. Store the Simulator in the carry
case provided until required for Re-calibration.
10. Offer up the Automatic Zero Point Assembly and tighten the 2 socket head screws.
11. Attach the 5 crimp connectors in the same orientation as previously noted.
12. Close the Transceiver casing and press the Maintenance key for 1 second until the LED
switches off. The instrument is now in ‘RUN’ mode.
The instrument is now ready for normal operation.