Model 4500 Premier Dust and Opacity Monitor
39
↵
↵↵
↵↵
→
←
↵
↵↵
↵↵
↑↓
↵
↵↵
↵↵
↑↓
↵
↵↵
↵↵
↑↓
↵
↵↵
↵↵
→
←
EXIT
Outputs 1:
↔
↵
↵↵
↵↵
Opacity
Outputs 1:
↑↓
↵
↵↵
↵↵
OD
Outputs 1:
↑↓
↵
↵↵
↵↵
Range:
1.6
Outputs 1:
↑↓
↵
↵↵
↵↵
Range:
3.0
Outputs 1:
↑↓
↵
↵↵
↵↵
During Cal:
TRACK
Outputs 1:
↑↓
↵
↵↵
↵↵
During Cal:
HOLD
Opacity:
0.0%
Output 1:
↑↓
↵
↵↵
↵↵
5.
Use the
←
and
→
keys to select the required option for Output
1 and then press the ‘ENTER’ (
↵
↵↵
↵↵
) key (notice no underlined
character). Press the
↑
and
↓
keys to scroll to the ‘Range’
option and press the ‘ENTER’ (
↵
↵↵
↵↵
) key again to display the
underlined character (The Constant Current option is used for
diagnostic purposes and is detailed in section B3-5).
6.
Press the
↑
and
↓
keys to select the maximum reading to be
allocated to the maximum output and then press the ‘ENTER’
(
↵
↵↵
↵↵
) key (notice no underlined character). Now press the
↑
and
↓
keys again to select the ‘During Cal’ option and press the
‘ENTER’ (
↵
↵↵
↵↵
) key.
7.
Using the
←
and
→
keys select ‘TRACK’ (to continue the output
signal during automatic calibration checks) or ‘HOLD’ (to retain
the last reading before the automatic calibration check until the
calibration is completed). Press the ‘ENTER’ (
↵
↵↵
↵↵
) key to set the
option (notice no underlined character) and then the press
‘EXIT’ key to return the display to ‘Output 1’.