Model 4500 Premier Dust and Opacity Monitor
49
5.
Use the
←
and
→
keys to select the required measurement
(Optical Density, Opacity or Dust) to be alarmed and then press
the ‘ENTER’ (
↵
) key (notice no underlined character). Press the
↑
and
↓
keys to scroll to the ‘AL1 trip’ (or ‘AL2 trip’ if setting
‘Alarm 2’) option and press the ‘ENTER’ (
↵
) key again (notice
the underlined character).
6.
Press the
↑
and
↓
keys to select the required level at which the
alarm is to trip and then press the ‘ENTER’ (
↵
) key (notice no
underlined character). Now press the
↑
and
↓
keys again to
select the ‘Delay’ option and press the ‘ENTER’ (
↵
) key (notice
the underlined character).
7.
Using the
↑
and
↓
keys select the required time period for the Alarm to
be delayed and press the ‘ENTER’ (
↵
) key (notice no underlined
character). Press the ‘EXIT’ key to return the display to ‘Alarm 1’.
8.
Press and hold the Maintenance key fro 1 second to return to ‘RUN’
mode (no LED illuminated).
The instrument alarms are now set.
↵
↵↵
↵↵
→
←
↵
↵↵
↵↵
↑↓
↵
↵↵
↵↵
↑↓
↵
↵↵
↵↵
↑↓
↵
↵↵
↵↵
↑↓
EXIT
Opacity:
0.0%
AL1 type:
Opacity
Opacity:
0.0%
AL1 type:
O.D.
Opacity:
0.0%
AL1 trip:
0.00
Opacity:
0.0%
AL1 trip:
15.00
Opacity:
0.0%
Delay:
5s
Opacity:
0.0%
Delay:
10s
Opacity:
0.0%
Alarm 1:
↑↓
↵
↵↵
↵↵